Add to Cart
The vertical laminar flow cabinet clean workbench is made up of indoor air filtered through a pre-filter, pressed into a static pressure tank by a centrifugal fan, and then filtered by a high-efficiency filter of air, and blown out from the air outlet of the air-efficient filter. Clean and uniform, it can eliminate the original air in the work area, and take away dust particles and biological particles to form a sterile and clean working environment.
It is widely used in scientific research and production departments such as medical and health, bio-pharmaceutical, food, medical science experiments, optics, electronics, laboratory testing, sterile microbial testing, plant tissue culture, etc., which require local clean and sterile working environment.
1).Body: 1.2mm thick high-quality cold-rolled steel plate, acid-washed and phosphating anti-corrosion, the surface is epoxy resin powder is not sprayed, anti-acid and anti-rust, countertop 304 stainless steel;
2).The single-sided vertical flow closes the countertop, effectively preventing the intrusion of outside air and the odor of the operating area from harming the human body;
3).LCD intelligent controller, the control is more accurate;
4).Can set sterilization delay time, sterilization pre-opening time, ambient temperature display, ambient temperature correction, date display and adjustment, sterilization lighting can be interlocked;
5).Can query the accumulated working time and remaining effective working time;
6).Five levels of wind speed are adjustable;
7).Has the function of replacing the high efficiency filter;
8).Sliding door: double-layer suspended glass door, open up and down, rubber cushion is embedded in the bottom of the sliding door, can be stopped at any position, convenient and flexible operation;
9).Provide anti-splash built-in power socket, can directly open the front cover, easy to replace the filter element, electronic components, etc., maintenance is more convenient;
Product Name | Single horizontal clean bench | Double horizontal clean bench | Double vertical clean bench |
Model | YN-VC-001 | YN-VC-002 | YN-VC-003 |
(W×D×H)mm External Dimension | 1000*700*1800 | 1200*700*1800 | 1500*700*1800 |
(W×D×H)mm Internal Work Zone | 900*650*600 | 1100*650*600 | 1400*650*600 |
Clean Class Level | 100Class 0.3µm(USA Federal Standard) | ||
Rated Air Flow | 1200m3/h | 1800m3/h | 2300m3/h |
Air Velocity | 0.3-0.6m/s | 0.3-0.6m/s | 0.3-0.6m/s |
HEPA Efficiency | 99.99% 0.3µm H13-H14 | ||
LUX | ≥300lux | ≥300lux | ≥300lux |
Power | 124W | 127W | 200W |
Noise | ≤65dB | ≤65dB | ≤65dB |
Power Supply | 220V/50Hz | 220V/50Hz | 220V/50Hz |
LED light | 9W*1 | 13W*1 | 18W*1 |
Material | Cabinet: epoxy powder coated steel or stainless steel, table-board: stainless steel |
Customized design is welcome.
Temperature: 5°C∽35°C;
Relative humidity: no more than 85% (+35°C);
Pressure: 86kPa∽106 kPa;
The maximum dust concentration: 3500 particles / L (dust particle size not less than 0.5μm, 1300);